A-Scan: Extending the Capabilities of EMAT Thickness Measurement

How to read an A-scan and when automatic thickness measurement is not enough
A-Scan: Extending the Capabilities of EMAT Thickness Measurement

The main operating mode of a thickness gauge is automatic measurement, when the screen shows a single number — the measurement result.
However, automatic measurement may not be sufficient when inspecting:
• complex-shaped objects,
• heavily corroded objects,
• heavily contaminated objects,
• cast components.

For experienced inspectors, an electromagnetic acoustic (EMAT) thickness gauge is a tool whose capabilities are significantly extended by the A-scan. Let us look at the EM1501 EMAT thickness gauge to see how to use the A-scan correctly and why it matters for reliable results.

Automatic measurements
The EM1501 EMAT thickness gauge measures metal thickness and displays the result in millimetres. For objects with simple geometry — a flat plate or a pipe — and light corrosion, this value is usually sufficient.
However, under heavy corrosion or contamination, on cast objects or complex-shaped components, this value alone is not enough for reliable measurements and conclusions about the object's condition. This is where the A-scan is required.

What does the A-scan show?

The A-scan is a plot with time (or distance, obtained by multiplying time by the speed of sound in the material) on the horizontal axis and the amplitude of the received signal, usually expressed in percent, on the vertical axis.
In the left part of the A-scan we see a very high-amplitude signal caused by the pulse generated by the thickness gauge. This area is called the dead zone, because echo signals cannot be received there.
After the dead zone comes an area where useful back-wall echoes are visible against the noise. For convenient analysis, the EM1501 EMAT thickness gauge can display both the real A-scan and its envelope.
If the transducer is too far from the test object — for example, because of a thick layer of rust, coating or other obstacles — the A-scan shows only noise to the right of the dead zone, without useful back-wall echoes.
As soon as the transducer is brought close enough to the test object to generate and receive the acoustic wave, back-wall echoes appear on the A-scan and the gauge starts measuring thickness from them. Therefore, when working on complex objects, the operator must monitor the level of back-wall echoes on the A-scan.
If the test object has parallel surfaces and is not heavily corroded, the A-scan shows smoothly decaying alternating echoes: the acoustic wave is generated at the object boundary, travels to the opposite wall, reflects, returns to the same point and continues bouncing between the walls until it decays.
Corrosion and defects on the A-scan


On heavily corroded objects the normal A-scan pattern is disrupted. Instead of a clear sequence of back-wall echoes, the A-scan becomes an almost random signal. Corrosion on the transducer side makes the wave radiate not along the normal but in other directions, from which waves never return; corrosion on the far side reflects the wave at an angle in various directions, and such reflections also never return to the transducer. In this case we can conclude that a defect or corrosion in the test object is interfering with the measurement, and slightly move the transducer on the surface in any direction to find a position where back-wall echoes are present without a strong interference pattern and the measurement is correct.
Geometry matters


Besides corrosion, the shape of the object can affect thickness measurement. When inspecting non-standard surfaces, such as areas with bends or sharp transitions, the wave may reflect at an angle and travel sideways without returning to the transducer. In such cases the A-scan helps identify areas where echo signals actually return, allowing a thickness measurement.
When automation is not enough: manual modes and gates



Automatic thickness measurement works well as long as the A-scan contains unambiguously interpretable echoes. However, in inhomogeneous material the automatic mode may work incorrectly due to ambiguous echoes on the A-scan. For such cases the EM1501 EMAT thickness gauge provides two manual algorithms — with one or two gates.
Each gate is a selected time interval in which the maximum echo amplitude is calculated.
Using two gates allows later echoes to be used for calculations when the first ones are weakened or noisy. For example, in the case above the first echo lies in the dead zone because the test object is thin. In some areas the third and fourth back-wall echoes drop almost to the noise level, so the gauge takes into account only one echo, treats it as the first (since subsequent echoes fell to the noise level) and calculates the thickness incorrectly, doubling the actual value. The manual two-gate mode allows the fifth and sixth echoes to be used as the basis and the thickness to be measured correctly.
This effect can be caused by metal anisotropy. By metal anisotropy we mean the dependence of the velocity of a polarized shear wave on the direction of oscillation. When anisotropy is present, a circularly polarized wave splits into two shear waves, each travelling at a slightly different velocity. Sometimes both waves can be seen on the A-scan.

Ignoring 'spurious' echoes


Sometimes certain echoes need to be ignored. For example, on the complex-shaped object shown (a rail), we see echoes at the beginning of the A-scan that came from a distance of 16–17 mm. The rail thickness is much greater, so if we need to measure the rail thickness, we explicitly set the interval in which we look for thickness by switching the gauge to the manual algorithm and placing one gate around the back-wall echo. The thickness gauge will only consider the echo in this zone and will not react to the 'spurious' echo that interferes with the measurement.

This method of ignoring some echoes is well suited for inspecting thick but heavily corroded objects, cast iron and other cast components, where spurious interfering echoes arise from the coarse-grained structure.


Improving measurement stability: signal accumulation
When noise is strong enough to cause unstable readings, another useful function helps — adjustment of the number of accumulations. Increasing the number of accumulations makes the EM1501 EMAT thickness gauge average several measurements, eliminating random noise. This slightly slows down the measurement but significantly improves reading stability.

A-scan as a decision-making tool
For experienced NDT specialists, the A-scan is an important tool for decision-making, accuracy and inspection reliability.
With its help a specialist can not only measure thickness but also understand why a particular value appeared on the screen and whether it can be trusted.
If you work with complex-shaped objects, heavily corroded objects or conditions where the automatic mode is not enough, use the capabilities of the A-scan. It is one of the tools that unlock the full potential of EMAT inspection.

Signal accumulation