Ultrasonic Thickness Gauge

up to 6 mm
Just bring the transducer to the surface of the object being tested
Works both on flat surfaces and on surfaces with a radius of curvature of 10 mm and more
For standard EMA transducers, the operating temperature range on the test object surface is -20…+80 °С. Special EMT14014T transducer measures thickness on objects heated up to 750 °C


The EM1401 UT allows the user to analyze A-scans, visualize B-scans (three display options), and save data conveniently on C-scans in a matrix form. If necessary, the user can work with gates, select a thickness measurement mode, signal processing parameters, etc.
The EM1401 UT can operate with:
• standard EMA transducers with a permanent magnet.
• piezoelectric transducers:
– transducer 5 MHz, aperture diameter is 10 mm,
– transducer 5 MHz, aperture diameter is 7.2 mm,
– transducer 10 MHz, aperture diameter is 5 mm.
The EM1401 UT implements a special algorithm that accounts for changes in the rate of sound propagation in metal induced by temperature.
Technical specifications
The EM1401 UT thickness gauge
-
Range of operating ambient temperature-20…+50 °С
-
Duration of continuous work without battery recharge7 hours
-
Range of sound velocity setting1000…9999 m/s with 1 m/s step
-
Highest number of measurements per second16
-
Dimensions232 x 135 x 44 mm
For EMA transducers:
-
Range of measured thickness for steel2…80 mm
-
Range of measured thickness for steel without clearance2…200 mm
-
Thickness measurement error 2…25 mm0.08+0.001·Н mm
-
Thickness measurement error 25…200 mm0.1+0.005·Н mm
-
Permissible gap between the EMT14012 transducer and the test objectUp to 4 mm
-
Permissible gap between the ЕМТ14013 transducer and the test objectUp to 6 mm
-
Permissible transducer tilt relative to the normal to the test object surface±25 °
-
Lowest permissible radius of curvature of the test object surface≥10 mm
-
Operating frequency of the device4 MHz
-
Range of operating temperature on the test object surface-20…+80 °C
(-20…+750 °C where the EMT14014Т transducer is used)
For piezoelectric transducers:
-
Range of measured thickness for steel0.5…300 mm
-
Thickness measurement error 0.5…25 mm0.08 mm
-
Thickness measurement error 25…300mm0.1 mm
-
Lowest permissible radius of curvature of the test object surface25 mm
-
Operating frequency of the device5 MHz, 10 MHz
-
Range of operating temperature on the test object surface-10…+60 °C
ACCESSORIES